Brief Evaluation of Olink Reveal Proximity Extension Assay for High-Throughput Proteomics: A Case Study Using NIST SRM 1950 and Two Spike-In Protein Standards

Olink Reveal 邻近延伸分析法在高通量蛋白质组学中的应用简述:以 NIST SRM 1950 和两种外源蛋白标准品为例的案例研究

阅读:1

Abstract

Plasma proteomics has regained attention in recent years through advancements in mass spectrometry instrumentation and sample preparation as well as new high-throughput affinity-based technologies. Here, we evaluate the analytical performance of the new Olink Reveal platform, a proximity extension assay (PEA)-based technology quantifying 1034 proteins and covering many biological pathways, in particular immune system processes. Using spiked-in recombinant Interleukin-10 (IL-10) and vascular endothelial growth factor D (VEGF-D) in the NIST SRM 1950 plasma standard, we assessed the linearity, sensitivity, precision, and accuracy of the Olink Reveal assay. The results demonstrated strong linear relationships (R(2) 0.922-0.953) for both IL-10 and VEGF-D across spiked-in concentrations, confirming the robust technical performance for these two proteins in the Olink Reveal platform. The resulting data contain no sensitive or personally identifiable information and are therefore suitable for use in benchmarking and software development. The data are publicly available in the PRIDE repository with identifier PAD000009.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。