Detection of mpox and other orthopoxviruses using a lateral flow device as a point-of-care diagnostic

利用侧向层析装置进行即时诊断,检测痘病毒和其他正痘病毒。

阅读:3

Abstract

In 2022, the World Health Organization declared the worldwide outbreak of mpox to be a public health emergency of international concern. The causative monkeypox virus (MPXV) belonged to clade IIb and is transmitted through sexual contact with a low case fatality rate (0.1%), which, together with under-detection, all contributed to a rapid global spread particularly within the MSM (men who have sex with men) community. As MPXV clade II remains circulating worldwide, a new outbreak of the more fatal clade I disease has been declared in Central and East Africa, and remains uncontrolled in part due to the lack of point-of-care (POC) diagnostics for rapid decisions on treatment and self-isolation. To address the lack of POC solutions for mpox, we have designed and evaluated an orthopoxvirus-specific lateral flow device (LFD) that could be used for the diagnosis of mpox. Using an LFD comprising four monoclonal antibodies against the A27 protein, we demonstrate sensitivity to 3 × 10(5) pfu/mL. This sensitivity is expected to be sufficient for the detection of MPXV from lesion sites and may also be sufficient for other sample types such as saliva and urine. We found that the presence of guanidinium thiocyanate, a common ingredient in inactivating viral transport media, masked the LFD antigen, resulting in false negatives. POC diagnosis of mpox may be possible using an LFD to reduce delays arising from sample shipment to centralized laboratory testing facilities. In order to achieve this, our work demonstrates that an LFD-optimized buffer is required, as the sample collection buffer may have a detrimental impact on sensitivity for clinical material.IMPORTANCEMpox cases have dramatically increased both in traditionally monkeypox virus endemic countries and also worldwide. This increase comes at a time when immunity derived from smallpox vaccination is no longer available. Diagnosis of mpox is complicated due to both disease presentation and the availability of local diagnostic laboratories. The availability of a point-of-care diagnostic tool such as an lateral flow device (LFD) would play an important role to both diagnose and prevent onward transmission. This manuscript provides developers and assessors with key data for defining true sensitivity and specificity of a successful LFD in addition to buffer conditions for sample collection.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。