Comparison of Four Systems for SARS-CoV-2 Antibody at Three Time Points after SARS-CoV-2 Vaccination

SARS-CoV-2疫苗接种后三个时间点四种SARS-CoV-2抗体检测系统的比较

阅读:1

Abstract

BACKGROUND: Immunity against severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) wanes over time after vaccination. METHODS: We compared SARS-CoV-2 antibody levels in serial samples from 350 vaccinated individuals at 3 time points (3 weeks after the first or second dose and before the third dose) with 4 assays: GenScript cPASS SARS-CoV-2 neutralization antibody detection kits (cPASS), Siemens SARS-CoV-2 IgG (sCOVG), Abbott SARS-CoV-2 IgG II Quant (CoV-2 IgG II), and an Immuno-On™ COVID-19 IgG test (Immuno-On IgG). Antibody levels by time, concordance between assays, and values from other tests corresponding to the percent inhibition results in cPASS were assessed. RESULTS: The median values at three time points were 49.31%, 90.87%, and 53.38% inhibition for cPASS, 5.39, 13.65, and 2.24 U/mL for sCOVG, 570.25, 1279.65, and 315.80 AU/mL for CoV-2 IgG II, and 223.22, 362.20, and 62.20 relative units (RU) for Immuno-On IgG. The concordance with cPASS at each time point ranged from 0.735 to 0.984, showing the highest concordance in the second sample and lowest concordance in the third in all comparative tests. The values corresponded to 30% inhibition, and the cutoffs of cPASS, were 2.02 U/mL, 258.6 AU/mL, and 74.2 RU for each test. Those for 50%, 70%, and 90% inhibition were 3.16, 5.66, and 8.26 U/mL for sCOVG, while they were 412.5, 596.9, and 1121.6 AU/mL for CoV-2 IgG II and 141.8, 248.92, and 327.14 RU for Immuno-On IgG. CONCLUSIONS: This study demonstrated the dynamic changes in antibody values at different time points using four test systems and is expected to provide useful baseline data for comparative studies and standardization efforts in the future.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。