Evaluation of the analytical performance of a point-of-care analyzer for the measurement of feline serum thyroxine concentration in comparison with a chemiluminescence enzyme immunoassay

评估即时分析仪测定猫血清甲状腺素浓度的分析性能,并与化学发光酶免疫测定法进行比较

阅读:1

Abstract

BACKGROUND: Total thyroxine (TT4) measurement is used to assess thyroid status in cats. OBJECTIVES: The aim of the prospective study was to evaluate the analytical performance of the point-of-care analyzer (POCA) Immuno AU10V using the v-T4 test kit for feline TT4 measurement. Additionally, method comparison with a benchtop analyzer (IMMULITE 2000) was done. METHODS: Validation included linearity, inter- and intra-assay precision, precision near the lower limit of quantification (LloQ), and interference testing for hemoglobin, lipid, and bilirubin. Correlation and bias were assessed. RESULTS: Linearity was given within the dynamic range. Coefficients of variation (CV) were ≤4% near the LloQ as well as for intra-and inter-assay precision. No interference was observed for lipid and bilirubin, while hemoglobin caused a negative bias of 28%. Method comparison included 74 samples within three TT4 concentration ranges (0.5-3.7, >3.7-5.13, >5.13-8 μg/dL). Correlation between POCA and reference method was excellent (r(s) = 0.95) with a slight proportional bias of 4.5%. TE(obs) was between 7.0% and 9.8%. Despite substantial agreement, discordant results on thyroid status occurred in 15% of samples. CONCLUSIONS: The analytical performance of the POCA was excellent, as was its correlation with the reference method. Except for the interferent effect of hemoglobin, the TE(obs) was 8 μg/dL, and de novo reference intervals need to be established.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。