Interfering lipoproteins in magnetic field-assisted agglutination of superparamagnetic particles immunoassay

干扰脂蛋白在磁场辅助超顺磁性颗粒凝集免疫测定中的作用

阅读:1

Abstract

OBJECTIVE: The technology of magnetic field-assisted immuno-agglutination of superparamagnetic particles allows sensitive detection of biomarkers in whole blood. However, we observed non-specific agglutination (NSA), due to interfering plasma proteins, that negatively affects C-reactive protein immunoassay. The objective of the study was to identify the plasma proteins involved and to eliminate these interferences. DESIGN AND METHODS: Plasma was fractionated by size exclusion HPLC and each fraction was tested for non-specific agglutination. In addition, plasma proteins bound to magnetic particles were analyzed by SDS-gel electrophoresis and identified by mass spectrometry. RESULTS: We found that NSA was due to the binding of some lipoproteins to the particles. NSA was observed in the presence of purified LDL and VLDL but not HDL. NSA was mediated by the binding of ApoB100 to magnetic particles through its heparin binding sites. These interferences could be eliminated by addition of heparin or other polyanions like dextran sulfate to the assay buffer. CONCLUSION: NSA results from the binding of some plasma lipoproteins to magnetic particles. The use of a polyanion to eliminate these interferences allows the formulation of a stable reagent.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。