Use of the Filovirus Animal Non-Clinical Group (FANG) Ebola virus immuno-assay requires fewer study participants to power a study than the Alpha Diagnostic International assay

与 Alpha Diagnostic International 的检测方法相比,使用丝状病毒动物非临床组 (FANG) 埃博拉病毒免疫测定法所需的受试者人数更少,即可满足研究的统计效力要求。

阅读:1

Abstract

As part of the scientific community's development of medical countermeasures against Ebola virus disease, optimization of standardized assays for product evaluation is paramount. The recent outbreak heightened awareness to the scarcity of available assays and limited information on performance and reproducibility. To evaluate the immunogenicity of vaccines entering Phase I-III trials and to identify survivors, two enzyme-linked immunosorbent assays, the Filovirus Animal Non-Clinical Group assay and the Alpha Diagnostics International assay, were evaluated for detection of immunoglobulin G against Ebola virus glycoprotein. We found that the Filovirus Animal Nonclinical Group assay produced a wider range of relative antibody concentrations, higher assay precision, larger relative accuracy range, and lower regional background. Additionally, to sufficiently power a vaccine trial, use of the Filovirus Animal Nonclinical Group assay would require one third the number of participants than the Alpha Diagnostics International assay. This reduction in needed study participants will require less money, fewer man hours, and much less time to evaluate vaccine immunogenicity.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。