A point-of-care microfluidic biochip for quantification of CD64 expression from whole blood for sepsis stratification

用于量化全血中 CD64 表达以进行脓毒症分层的即时护理微流控生物芯片

阅读:5
作者:U Hassan, T Ghonge, B Reddy Jr, M Patel, M Rappleye, I Taneja, A Tanna, R Healey, N Manusry, Z Price, T Jensen, J Berger, A Hasnain, E Flaugher, S Liu, B Davis, J Kumar, K White, R Bashir

Abstract

Sepsis, a potentially life-threatening complication of an infection, has the highest burden of death and medical expenses in hospitals worldwide. Leukocyte count and CD64 expression on neutrophils (nCD64) are known to correlate strongly with improved sensitivity and specificity of sepsis diagnosis at its onset. A major challenge is the lack of a rapid and accurate point-of-care (PoC) device that can perform these measurements from a minute blood sample. Here, we report a PoC microfluidic biochip to enumerate leukocytes and quantify nCD64 levels from 10 μl of whole blood without any manual processing. Biochip measurements have shown excellent correlation with the results from flow cytometer. In clinical studies, we have used PoC biochip to monitor leukocyte counts and nCD64 levels from patients' blood at different times of their stay in the hospital. Furthermore, we have shown the biochip's utility for improved sepsis diagnosis by combining these measurements with electronic medical record (EMR).