Charge-Sensitive Optical Detection of Small Molecule Binding Kinetics in Normal Ionic Strength Buffer

正常离子强度缓冲液中小分子结合动力学的电荷敏感光学检测

阅读:10
作者:Runli Liang, Guangzhong Ma, Wenwen Jing, Yan Wang, Yunze Yang, Nongjian Tao, Shaopeng Wang

Abstract

Most label-free detection technologies detect the masses of molecules, and their sensitivities thus decrease with molecular weight, making it challenging to detect small molecules. To address this need, we have developed a charge-sensitive optical detection (CSOD) technique, which detects the charge rather than the mass of a molecule with an optical fiber. However, the effective charge of a molecule decreases with the buffer ionic strength. For this reason, the previous CSOD works with diluted buffers, which could affect the measured molecular binding kinetics. Here, we show a technique capable of detecting molecular binding kinetics in normal ionic strength buffers. An H-shaped sample well was developed to increase the current density at the sensing area to compensate the signal loss due to ionic screening at normal ionic strength buffer, while keeping the current density low at the electrodes to minimize the electrode reaction. In addition, agarose gels were used to cover the electrodes to prevent electrode reaction generated bubbles from entering the sensing area. With this new design, we have measured the binding kinetics between G-protein-coupled receptors (GPCRs) and their small molecule ligands in normal buffer. We found that the affinities measured in normal buffer are stronger than those measured in diluted buffer, likely due to the stronger electrostatic repulsion force between the same charged ligands and receptors in the diluted buffer.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。