BubbleDrive, a low-volume incubation chamber for acute brain slices

BubbleDrive,一种用于急性脑切片的低容量孵化室

阅读:12
作者:Aditi Naik #, Vidar Jensen #, Cecilie Bugge Bakketun, Rune Enger, Sabina Hrabetova, Jan Hrabe

Abstract

Acute brain slices are a common and useful preparation in experimental neuroscience. A wide range of incubation chambers for brain slices exists but only a few are designed with very low volumes of the bath solution in mind. Such chambers are necessary when high-cost chemicals are to be added to the solution or when small amounts of substances released by the slice are to be collected for analysis. The principal challenge in designing a very low-volume incubation chamber is maintaining good oxygenation and flow without mechanically disturbing or damaging the slices. We designed and validated BubbleDrive, a 3D-printed incubation chamber with a minimum volume of 1.5 mL which can hold up to three coronal mouse slices from one hemisphere. It employs the carbogen gas bubbles to drive the flow circulation in a consistent and reproducible manner, and without disturbing the brain slices. The BubbleDrive design and construction were successfully validated by comparison to a conventional large-volume incubation chamber in several experimental designs involving measurements of extracellular diffusion parameters, the electrophysiology of neuronal and astrocytic networks, and the effectiveness of slice incubation with hyaluronidase enzyme.