Background
Biodegradable suture anchors are associated with higher redislocation rates. This study examined whether the biocompatibility and/or biomechanical properties of suture anchors contribute to the increase in complications.
Conclusion
BioKnotless and Suretac II anchors are cytotoxic. The BioKnotless biodegradable anchor has significantly lower failure load. Absorbable suture anchors may cause higher redislocation of arthroscopic Bankart repair.
Methods
Human glenohumeral capsule cells were cultured with 4 types of suture anchors, Opus LabraFix (titanium alloy; ArthroCare, Austin, TX, USA), PushLock (poly-ether-ether-ketone; Arthrex, Naples, FL, USA), BioKnotless (poly-l-lactic acid; DePuy Mitek, Warsaw, IN, USA), and Suretac II (polyglycolic acid; Smith & Nephew, London, UK), to measure cell viability and pH. Four groups of 6 ovine shoulders were used to repair the labrum, which was completely detached from the glenoid rim anteroinferiorly and reattached with 2 suture anchors and subject to failure load testing.
Results
In cell culture, BioKnotless at 48 and 72 hours (85.2% ± 2.1% and 84.5% ± 3.6%) and Suretac II groups (33.9% ± 3.1% and 42.8% ± 6.4%) had fewer viable cells compared with control (P = .048). The pH of Suretac II was lower than control (7.51 to 7.65) at 24 hours (7.31 ± 0.08, P = .049), 48 hours (7.25 ± 0.02, P = .046), and 72 hours (7.29 ± 0.04, P = .04). During mechanical testing, 83% of repairs failed by the capsule tearing. Among the anchors, the BioKnotless repair group had a significantly lower failure load (37 ± 5 N) compared with the PushLock (61 ± 7 N), Opus (60 ± 6 N), and Suretac II (57 ± 7 N) groups (P = .038).