On-chip solid-phase extraction pre-concentration/focusing substrates coupled to atmospheric pressure matrix-assisted laser desorption/ionization ion trap mass spectrometry for high sensitivity biomolecule analysis

用于高灵敏度生物分子分析的芯片上固相萃取预浓缩/聚焦基底与大气压基质辅助激光解吸/电离离子阱质谱联用

阅读:1

Abstract

Atmospheric pressure matrix-assisted laser desorption/ionization (AP-MALDI) has proven a convenient and rapid method for ion production in the mass spectrometric (MS) analysis of biomolecules. AP-MALDI and electrospray ionization (ESI) sources are easily interchangeable in most mass spectrometers. However, AP-MALDI suffers from less-than-optimal sensitivity due to ion losses during transport from the atmosphere into the vacuum of the mass spectrometer. Here, we study the signal-to-noise ratio (S/N) gains observed when an on-chip dynamic pre-concentration/focusing approach is coupled to AP-MALDI for the MS analysis of neuropeptides and protein digests. It was found that, in comparison with conventional AP-MALDI targets, focusing targets showed (1) a sensitivity enhancement of approximately two orders of magnitude with S/N gains of 200-900 for hydrophobic substrates, and 150-400 for weak cation-exchange (WCX) substrates; (2) improved detection limits as low as 5 fmol/microL for standard peptides; (3) significantly reduced matrix background; and (4) higher inter-day reproducibility. The improved sensitivity allowed successful tandem mass spectrometric (MS/MS) sequencing of dilute solutions of a derivatized tryptic digest of a protein standard, and enabled the first reported AP-MALDI MS detection of neuropeptides from Aedes aegypti mosquito heads.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。