Comparison of a Traditional and Novel Evoked Compound Action Potentials Recording Approach and Evoked Auditory Brainstem Responses in Pediatric Cochlear Implants Users

比较传统和新型诱发复合动作电位记录方法以及儿童人工耳蜗植入者诱发听觉脑干反应

阅读:1

Abstract

OBJECTIVES: Electrically evoked compound action potentials (eCAP) recordings are widely used in functional evaluation and fitting of cochlear implants (CI) in clinics. We compared the results from two eCAP recording approaches (StandardART and FineGrain, MED-EL, Austria). The FineGrain method is more advanced than the Auditory Nerve Response Telemetry (StandardART) method in terms of the stimulation and algorithm for the eCAP threshold detection. To understand the benefits of these alterations, we compared the two methods on a larger scale in pediatric CI users alongside evoked auditory brainstem responses (eABR). MATERIALS AND METHODS: We collected the eCAP recordings obtained with both methods from a population of pediatric subjects with CI, either intra- or post-operatively. The eABR recordings were only collected post-operatively. For comparability reasons, we used the same stimulation rate and similar amplitude levels for all three approaches. RESULTS: Our results demonstrate that, although the success rates are similar, the FineGrain method outperforms traditional StandardART in terms of robustness and measurement duration. The eCAP recordings in general outperform the eABR in terms of speed. CONCLUSION: We conclude that the eCAP recordings are the method of choice for measuring the auditory neural activity, and FineGrain outperforms StandardART. From the three investigated approaches, we conclude that FineGrain performed best and should be the first-choice method in pediatric patients.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。