Calibration Fluids and Calibration Equations: How Choices May Affect the Results of Density Measurements Made with U-Tube Densimeters

校准液和校准方程:选择如何影响U型管密度计的密度测量结果

阅读:1

Abstract

Data on the calibration fluids water and toluene individually, and the combination of those two sets of data, were correlated with two different equations (resulting in six sets of calibration equation parameters) to analyze data measured with a vibrating-tube densimeter on the lubricants pentaerythritol tetrapentanoate (POE5), pentaerythritol tetraheptanoate (POE7), and pentaerythritol tetranonanoate (POE9) at temperatures from 270 K to 470 K and pressures from 0.5 MPa to 50 MPa. The objective was to explore how the calculated densities of the lubricants would differ based on the calibration equations and calibration fluid(s). The viscosities of the measured lubricants are much greater than those of the calibration fluids, and because there has long been a question of how measurements of higher viscosity fluids are affected when measured with a vibrating-tube densimeter, combinations of calibration fluid(s) and equations were tested to explore the role they play in obtaining accurate results. For the lubricants studied herein, more accurate results were obtained with a calibration fitted to multiple calibration fluids, while the consistency of results was more equation dependent.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。