Results of a CIE Detector Response Intercomparison

CIE探测器响应比对结果

阅读:1

Abstract

A total of fifteen laboratories participated in the CIE detector response intercomparison which was designed to assess the level of agreement among participating laboratories in the absolute measurement (with respect to SI) of photodetector response in the visible spectral region. Most participants were either commercial laboratories or university laboratories with the National Institute of Standards and Technology (NIST) serving as the host laboratory. Each laboratory determined the absolute response of each of two silicon photodiode radiometers which were designed for the intercomparison by NIST. Approximately two-thirds of the laboratories reported response values which agreed with the NIST values to within ±1.0% at the two wavelengths of 488 and 633 nm.