Statistical Analysis of a Round-Robin Measurement Survey of Two Candidate Materials for a Seebeck Coefficient Standard Reference Material

对两种塞贝克系数标准参考物质候选材料进行循环测量调查的统计分析

阅读:1

Abstract

In an effort to develop a Standard Reference Material (SRM™) for Seebeck coefficient, we have conducted a round-robin measurement survey of two candidate materials-undoped Bi2Te3 and Constantan (55 % Cu and 45 % Ni alloy). Measurements were performed in two rounds by twelve laboratories involved in active thermoelectric research using a number of different commercial and custom-built measurement systems and techniques. In this paper we report the detailed statistical analyses on the interlaboratory measurement results and the statistical methodology for analysis of irregularly sampled measurement curves in the interlaboratory study setting. Based on these results, we have selected Bi2Te3 as the prototype standard material. Once available, this SRM will be useful for future interlaboratory data comparison and instrument calibrations.