Optimal flow of high-flow nasal cannula oxygenation to prevent desaturation during sedation for bronchoscopy: a randomized controlled study

高流量鼻导管氧疗的最佳流量可预防支气管镜检查镇静期间的氧饱和度下降:一项随机对照研究

阅读:1

Abstract

BACKGROUND: Although high-flow nasal cannula (HFNC) oxygenation is currently recommended to prevent desaturation during sedation for bronchoscopy, there is no consensus on an optimal flow rate. OBJECTIVE: To determine the optimal oxygen flow rate for HFNC to effectively prevent desaturation during sedation for bronchoscopy. DESIGN: Prospective, randomized, and controlled study. METHODS: Patients (n = 240) scheduled for bronchoscopy were randomized to receive HFNC with propofol sedation (fraction of inspired oxygen, 100%) at one of six flow rates of 10, 20, 30, 40, 50, and 60 L/min, designated as groups 1-6, respectively. RESULTS: The incidence of desaturation significantly decreased by increasing the oxygen flow rate (42.5%, 17.5%, 15%, 10%, 2.5%, and 0% for groups 1-6, respectively, p < 0.0001). The optimal oxygen flow rate for HFNC determined by probit regression to effectively prevent desaturation in 95% of patients was 43.20 (95% confidence interval, 36.43-55.96) L/min. The requirement for airway intervention was significantly decreased by increasing the oxygen flow rate. CONCLUSION: An HFNC flow rate of 50-60 L/min is recommended to prevent desaturation during sedation for bronchoscopy. REGISTRATION: NCT05298319 at ClinicalTrials.gov.