Comparison of ejection fraction calculation between CT and SPECT at high heart rate: A dynamic cardiac phantom study

高心率下CT与SPECT射血分数计算方法的比较:一项动态心脏模型研究

阅读:1

Abstract

The purpose of this study is to compare the ejection fraction (EF) calculation of CT and SPECT at high heart rate. A dynamic cardiac phantom with programmable end-systolic volume (ESV), end-diastolic volume (EDV), and heart rate was used to compare CT, which has high spatial resolution (< 1 mm) and modest temporal resolution of 175 msec, and SPECT, which has high temporal resolution of 16 bins per cardiac cycle but poor spatial resolution (> 1 cm) in EF, ESV, and EDV at the heart rates ≤ 100 bpm for EF = 30 (disease state) and EF = 60 (healthy state). EF calculations for SPECT were accurate in 2% for 40 to 100 bpm for both EF = 30 and EF = 60, and were not heart rate dependent although both ESV and EDV could be underestimated by 18-20%. EF calculations for CT were accurate in 2.2% for 40 and 60 bpm. Inaccuracy in EF calculations, ESV and EDV estimates increased when the heart rate or EF increased. SPECT was accurate for EF calculation for the heart rates ≤ 100 bpm and CT was accurate for the heart rates of ≤ 60 bpm. CT was less accurate for the high heart rates of 80 and 100 bpm, or high EF = 60.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。