PET/CT imaging: effect of respiratory motion on apparent myocardial uptake

PET/CT成像:呼吸运动对心肌表观摄取的影响

阅读:1

Abstract

BACKGROUND: Positron emission tomography (PET) attenuation correction (AC) using computed tomography (CT) can be affected by respiratory motion: hi-speed CT captures 1 point of the respiratory cycle while PET emission data averages many cycles. We quantified the changes in apparent myocardial uptake due to this respiratory-induced CT attenuation mismatch. METHODS: Twenty-two patients undergoing fluorine-18 fluorodeoxyglucose (FDG) PET/CT received 3 sequential CT scans at normal resting end-inspiration (CT(INSPIR)), ending expiration (CT(EXPIR)), and at midvolume between end-expiration and end-inspiration (CT(MIDVOL)). A pneumotachometer measured absolute changes in lung volume. Seven subjects also underwent a 3-minute transmission scan with a 68Ge rotating rod source (RRS). The PET emission data set was reconstructed up to 4 times using CT(EXPIR), CT(INSPIR), CT(MIDVOL), and RRS AC maps. Relative heart position and cardiac uptake was measured for each CT attenuation correction. RESULTS: Respiratory motion produced marked changes in global and regional myocardial uptake. Changes were large in the lateral and anterior regions at the lung-soft tissue interface (up to 30% using CT(INSPIR) compared to CT(EXPIR) for AC) and smaller in the septal region (10% or less). Data corrected with CT(EXPIR) agreed best with the RRS. CONCLUSION: Respiratory effects can introduce large inhomogeneities in apparent myocardial uptake when CT is used for attenuation correction.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。