Validation of a point-of-care handheld blood total calcium analyzer in postpartum dairy cows

验证即时手持式血液总钙分析仪在产后奶牛中的应用

阅读:1

Abstract

Our objective was to validate a point-of-care handheld blood total calcium analyzer (Ca meter, CM; TD-5220 Vet Ca(2+), TaiDoc, New Taipei, Taiwan) to estimate circulating Ca concentrations in postpartum dairy cows. Whole blood was collected from 251 multiparous cows between 1 and 4 d in milk from 2 commercial dairy herds in Ontario, Canada. Blood total calcium concentration (tCa) was analyzed in whole blood, fresh plasma, and thawed plasma, and compared with tCa results from thawed serum analyzed in a diagnostic laboratory (using a Cobas Calcium Gen 2 kit, Roche Diagnostics, Indianapolis, IN) as the reference test (RT). Lin's concordance correlation coefficient (βrho;) and Bland-Altman (B-A) plots were assessed to evaluate the agreement between the RT and CM results in each type of sample. Receiver operating characteristic curve analyses were used to describe the accuracy of each test against the categorized RT results (at a cut-point of ≤2.14 mmol/L). Samples where the meter gave a nonquantitative result ("high" or "low"; thawed plasma: 3/247; fresh plasma: 6/100; and whole blood: 20/98) were not included in the βrho; and B-A analyses. Lin's correlation coefficients demonstrated poor agreement between tests (thawed plasma: βrho; = 0.16; fresh plasma: βrho; = 0.21; and whole blood: βrho; = 0.23). Fresh plasma (using a cut-point of 2.55 mmol/L as measured on the CM) had the greatest diagnostic sensitivity (72%), specificity (86%), and accuracy (77%) for determining subclinical hypocalcemia, but that would still misclassify 23% of samples. In addition to substantial variability, the B-A plots revealed bias with changing concentrations of calcium. Because of low sensitivity on whole blood (58%) or thawed plasma (56%), measurement with the CM is not recommended on these types of samples. This rapid and low-cost meter was not sufficiently accurate to quantify blood Ca concentration, but when used with fresh plasma it might be useful as a screening tool for subclinical hypocalcemia.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。