Optical microangiography provides depth-resolved images of directional ocular blood perfusion in posterior eye segment

光学微血管造影术可提供眼后段定向眼部血液灌注的深度分辨图像

阅读:1

Abstract

In this paper, we demonstrate that the optical microangiography (OAMG) is capable of depth-resolved imaging of directional blood perfusion within both retinal and choroid in the posterior segment of human eye. The study uses an OMAG system operating at 840 nm with an imaging speed at 27,000 A-scans per second. Sequentially registered multiple OMAG projection maps of small areas ( approximately 1 mm x 1 mm) are combined to provide directional blood flow images for a larger field of view. It takes approximately 3.7 sec to image a small area (1 x 1 mm(2)), and approximately 2.5 min for a larger field (3 x 3 mm(2)). Finally, we show superior performance of OMAG in providing functional images of capillary level microcirculation at different depths of retina and choroid that correlate well with the standard retinal pathology.