Fluorescence instrument response standards in two-photon time-resolved spectroscopy

双光子时间分辨光谱中的荧光仪器响应标准

阅读:5
作者:Rafal Luchowski, Mariusz Szabelski, Pabak Sarkar, Elisa Apicella, Krishna Midde, Sangram Raut, Julian Borejdo, Zygmunt Gryczynski, Ignacy Gryczynski

Abstract

We studied the fluorescence properties of several potential picosecond lifetime standards suitable for two-photon excitation from a Ti:sapphire femtosecond laser. The fluorescence emission of the selected fluorophores (rose bengal, pyridine 1, and LDS 798) covered the visible to near-infrared wavelength range from 550 to 850 nm. We suggest that these compounds can be used to measure the appropriate instrument response functions needed for accurate deconvolution of fluorescence lifetime data. Lifetime measurements with multiphoton excitation that use scatterers as a reference may fail to properly resolve fluorescence intensity decays. This is because of the different sensitivities of photodetectors in different spectral regions. Also, detectors often lose sensitivity in the near-infrared region. We demonstrate that the proposed references allow a proper reconvolution of measured lifetimes. We believe that picosecond lifetime standards for two-photon excitation will find broad applications in multiphoton spectroscopy and in fluorescence lifetime imaging microscopy (FLIM).