Impact of G29179T mutation on two commercial PCR assays for SARS-CoV-2 detection

G29179T突变对两种商业化SARS-CoV-2检测PCR方法的影响

阅读:2

Abstract

Nucleic acid amplification test (NAAT) is the gold standard for severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) detection. However, genetic mutations in the virus can affect the result. Cycle threshold (Ct) values of N genes and their association with mutations using SARS-CoV-2 positive specimens diagnosed by the Xpert Xpress SARS-CoV-2 were examined in this study. In total, 196 nasopharyngeal swab specimens were tested for SARS-CoV-2 infection using the Xpert Xpress SARS-CoV-2, and 34 were positive. WGS was performed for four outlier samples with increased ΔCt identified by Scatterplot analysis as well as seven control samples without increased ΔCt in the Xpert Xpress SARS-CoV-2. The presence of the G29179T mutation was identified as a cause of increased ΔCt. PCR using the Allplex™ SARS-CoV-2 Assay did not show a similar increase in ΔCt. Previous reports focusing on N-gene mutations and their effects on SARS-CoV-2 testing including the Xpert Xpress SARS-CoV-2 were also summarized. While a single mutation that impacts one target of a multiplex NAAT is not a true detection failure, mutation compromising NAAT target region can cause confusion of the results and render the assay susceptible to diagnostic failure.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。