Performance evaluation of automated chemiluminescence immunoassay based antigen detection - Moving towards more reliable ways to predict SARS-CoV-2 infection

基于自动化化学发光免疫分析的抗原检测性能评估——迈向更可靠的SARS-CoV-2感染预测方法

阅读:1

Abstract

Real-time reverse transcription- polymerase chain reaction (RT-PCR) has been the most reliable armoury for the diagnosis of COVID-19, considered to be the reference standard but fails to reproduce the correct predictability about the infectivity of the disease every time. Antigen detection however puts foothold in this aspect even though lacks in sensitivity, especially conventional Rapid Antigen Tests (RATs). Recently developed Chemiluminescence Immunoassay (CLIA) based antigen detection tests are promising and displayed better sensitivity. In the current study we have evaluated VITROS® SARS-CoV-2 Ag Test CLIA Kit, which was tested on 148 patient's samples attended to a tertiary care centre for testing of SARS-CoV-2. The performance of the kit was evaluated in comparison to RT-PCR and RAT and found to be a good test for antigen detection, best within the first few days of infection. The test has shown sensitivity of 94.3 % and specificity of 100 % in samples with corresponding Ct values of ≤25 by RT-PCR, which corresponds to high viral load and can predict ability of spreading the disease by the patients. With the results being semiquantitative along with improved sensitivity it can replace RATs for antigen detection for screening, provided good laboratory set up is included under consideration.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。