Analysis of external quality assessment samples revealed crucial performance differences between commercial RT-PCR assays for SARS-CoV-2 detection when taking extraction methods and real-time-PCR instruments into account

对外部质量评估样本的分析表明,在考虑提取方法和实时PCR仪器的情况下,用于检测SARS-CoV-2的商业化RT-PCR检测方法之间存在显著的性能差异。

阅读:2

Abstract

In limelight of the ongoing pandemic SARS-CoV-2 testing is critical for the diagnosis of infected patients, contact-tracing and mitigating the transmission. Diagnostic laboratories are expected to provide appropriate testing with maximum accuracy. Real-time reverse transcriptase PCR (RT-PCR) is the diagnostic standard. However, only a handful of studies have reviewed their performance in clinical settings. The aim of this study was to compare the performance of the overall analytical matrix including the extraction kit (BD MAX, Promega, Qiagen), the PCR instrument (Agilent Mx3005 P, BD MAX, Qiagen Rotor-Gene, Roche Cobas z 480) and the RT-PCR assay (Altona Diagnostics, CerTest Biotec, R-Biopharm AG) using predefined samples from proficiency testing organizers. The greatest difference of the cycle threshold values between the matrices was nine cycles. One borderline sample could not be detected by three out of twelve analytical matrices and yielded a false negative result. We therefore conclude that diagnostic laboratories should take the complete analytical matrix in addition to the performance values published by the manufacturer for a respective RT-PCR kit into account. With limited resources laboratories have to validate a wide range of kits to determine appropriate analytical matrices for detecting SARS-CoV-2 reliably. The interpretation of clinical results has to be adapted accordingly.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。