Comparison of continuous temperature measurement methods in the intensive care unit: standard bladder catheter measurements versus non-invasive transcutaneous sensors

重症监护病房中连续体温测量方法的比较:标准膀胱导管测量与无创经皮传感器测量

阅读:1

Abstract

The purpose of this study was to compare a wearable system for body core temperature measurement versus bladder and tympanic thermometers in an intensive care setting. The question was, if continuous non-invasive sensors in the intensive care unit represent an alternative to current standard methods of invasive continuous bladder temperature measurement methods?Between May and September 2023, a comparative investigation involving 112 patients was conducted in a 20-bed surgical intensive care unit to assess various temperature probes, including those placed in the tympanic tube, bladder, and skin. To achieve this, a wireless non-invasive sensor system provided by greenTEG AG, Switzerland, was affixed to different body locations (clavicular and lateral chest) of each catheterized patient (equipped with a temperature probe) admitted to the intensive care unit. Furthermore, tympanic temperatures were recorded at specified intervals. The measurement duration ranged from a minimum of six hours to a maximum of six days, resulting in the analysis of a total of 355 simultaneous temperature measurements.In this study, a wearable temperature measurement system attached to two different body sites revealed a consistent negative bias compared to bladder temperature. In addition, the measurements were particularly influenced by body constitution. The tested system in all patients showed a mean absolute error (MAE) of 0.45 °C for the lateral chest and 0.50 °C for the clavicular position. Tympanic measurements had a mean absolute error of 0.35 °C. In patients with body mass index (BMI) ≥ 25 the MAE increased to 0.5 °C for the lateral chest and 0.56 °C for the clavicular position. In contrast, the tympanic measurement had a reduced MAE of 0.32 °C, which is well below this threshold when compared to bladder measurements.In conclusion the investigated system did not meet the clinically relevant acceptance criteria and showed low precision in correctly identifying fever episodes compared to invasive temperature probes, however its main advantage lies in its continuity and non-invasiveness. This makes it a potential alternative to intermittent tympanic measurement devices. In this study we were able to show, that in at least one subset of patients, the non-invasive and continuous device demonstrated a precision comparable to tympanic measurements.The accuracy of all non-invasive methods was lower than in previous studies, suggesting that the use of bladder temperature as reference and user related variations may have introduced additional errors.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。