Accuracy of calculating mechanical power of ventilation by one commonly used equation

用一种常用公式计算通风机械功率的准确性

阅读:1

Abstract

Gattinoni's equation, [Formula: see text], now commonly used to calculate the mechanical power (MP) of ventilation. However, it calculates only inspiratory MP. In addition, the inclusion of PEEP in Gattinoni's equation raises debate because PEEP does not produce net displacement or contribute to MP. Measuring the area within the pressure-volume loop accurately reflects the MP received in a whole ventilation cycle and the MP thus obtained is not influenced by PEEP. The MP of 25 invasively ventilated patients were calculated by Gattinoni's equation and measured by integration of the areas within the pressure-volume loops of the ventilation cycles. The MP obtained from both methods were compared. The effects of PEEPs on MP were also evaluated. We found that the MP obtained from both methods were correlated by R(2) = 0.75 and 0.66 at PEEP 5 and 10 cmH(2)O, respectively. The biases of the two methods were 3.13 (2.03 to 4.23) J/min (P < 0.0001) and - 1.23 (- 2.22 to - 0.24) J/min (P = 0.02) at PEEP 5 and 10 cmH(2)O, respectively. These P values suggested that both methods were significantly incongruent. When the tidal volume used was 6 ml/Kg, the MP by Gattinoni's equation at PEEP 5 and 10 cmH(2)O were significantly different (4.51 vs 7.21 J/min, P < 0.001), but the MP by PV loop area was not influenced by PEEPs (6.46 vs 6.47 J/min, P = 0.331). Similar results were observed across all tidal volumes. We conclude that the Gattinoni's equation is not accurate in calculating the MP of a whole ventilatory cycle and is significantly influenced by PEEP, which theoretically does not contribute to MP.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。