True intratracheal oxygen concentration delivered by SentriO Oxy™ masks under various respiratory conditions: a bench study

SentriO Oxy™ 面罩在不同呼吸条件下输送的真实气管内氧浓度:一项体外研究

阅读:1

Abstract

SentriO Oxy™ is a newly available, Food and Drug Administration-approved oxygenation mask system that provides high oxygenation, even on low-flow (5-10 L/min) oxygen. This study aimed to accurately measure the intratracheal fraction of inspired oxygen (FiO(2)) using SentriO Oxy™ masks under relatively low oxygen flow rates. A manikin-ventilator-test lung simulation system was used. We measured FiO(2) at the level of the carina, 5 minutes after applying 45 different respiratory parameter combinations using SentriO Oxy™ masks. Tidal volume (TV) was set to 300, 500, and 700 mL; respiratory rate (RR) was set to 8, 12, 16, 20, and 24 breaths per minute; and oxygen flow rate was set to 6, 8, and 10 L/min. Our hypothesis was that FiO(2) would be proportional to the difference between oxygen flow rate and minute ventilation. FiO(2) measured by smaller TV, lower RR, or higher oxygen flows revealed a significantly higher value, confirming our hypothesis. In addition, using linear regression analysis, we found that TV, RR, and oxygen flow were all significant factors influencing the measured FiO(2). Our experiment proposed two prediction equations considering the oxygen flow rate, TV, and RR. The results of our study may provide information and prediction of FiO(2) for clinicians to use SentriO Oxy™ masks during sedative anesthetic procedures under low oxygen flow rates.