Effectiveness of consumer-grade contactless vital signs monitors: a systematic review and meta-analysis

消费级非接触式生命体征监测仪的有效性:系统评价和荟萃分析

阅读:1

Abstract

The objective of this systematic review and meta-analysis was to analyze the effectiveness of contactless vital sign monitors that utilize a consumer-friendly camera versus medical grade instruments. A multiple database search was conducted from inception to September 2020. Inclusion criteria were as follows: studies that used a consumer-grade camera (smartphone/webcam) to examine contactless vital signs in adults; evaluated the non-contact device against a reference medical device; and used the participants' face for measurement. Twenty-six studies were included in the review of which 16 were included in Pearson's correlation and 14 studies were included in the Bland-Altman meta-analysis. Twenty-two studies measured heart rate (HR) (92%), three measured blood pressure (BP) (12%), and respiratory rate (RR) (12%). No study examined blood oxygen saturation (SpO(2)). Most studies had a small sample size (≤ 30 participants) and were performed in a laboratory setting. Our meta-analysis found that consumer-grade contactless vital sign monitors were accurate in comparison to a medical device in measuring HR. Current contactless monitors have limitations such as motion, poor lighting, and lack of automatic face tracking. Currently available consumer-friendly contactless monitors measure HR accurately compared to standard medical devices. More studies are needed to assess the accuracy of contactless BP and RR monitors. Implementation of contactless vital sign monitors for clinical use will require validation in a larger population, in a clinical setting, and expanded to encompass other vital signs including BP, RR, and SpO(2).

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。