Lightweight physiologic sensor performance during pre-hospital care delivered by ambulance clinicians

轻型生理传感器在救护车医护人员提供的院前护理中的性能

阅读:1

Abstract

The aim of this study was to explore the impact of motion generated by ambulance patient management on the performance of two lightweight physiologic sensors. Two physiologic sensors were applied to pre-hospital patients. The first was the Contec Medical Systems CMS50FW finger pulse oximeter, monitoring heart rate (HR) and blood oxygen saturation (SpO2). The second was the RESpeck respiratory rate (RR) sensor, which was wireless-enabled with a Bluetooth(®) Low Energy protocol. Sensor data were recorded from 16 pre-hospital patients, who were monitored for 21.2 ± 9.8 min, on average. Some form of error was identified on almost every HR and SpO2 trace. However, the mean proportion of each trace exhibiting error was <10 % (range <1-50 % for individual patients). There appeared to be no overt impact of the gross motion associated with road ambulance transit on the incidence of HR or SpO2 error. The RESpeck RR sensor delivered an average of 4.2 (±2.2) validated breaths per minute, but did not produce any validated breaths during the gross motion of ambulance transit as its pre-defined motion threshold was exceeded. However, this was many more data points than could be achieved using traditional manual assessment of RR. Error was identified on a majority of pre-hospital physiologic signals, which emphasised the need to ensure consistent sensor attachment in this unstable and unpredictable environment, and in developing intelligent methods of screening out such error.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。