Impact of changes in systemic vascular resistance on a novel non-invasive continuous cardiac output measurement system based on pulse wave transit time: a report of two cases

系统血管阻力变化对基于脉搏波传导时间的新型无创连续心输出量测量系统的影响:两例病例报告

阅读:1

Abstract

The inaccuracy of arterial waveform analysis for measuring continuos cardiac output (CCO) associated with changes in systemic vascular resistance (SVR) has been well documented. A new non-invasive continuous cardiac output monitoring system (esCCO) mainly utilizing pulse wave transit time (PWTT) in place of arterial waveform analysis has been developed. However, the trending ability of esCCO to measure cardiac output during changes in SVR remains unclear. After a previous multicenter study on esCCO measurement, we retrospectively identified two cases in which apparent changes in SVR developed in a short period during data collection. In each case, the trending ability of esCCO to measure cardiac output and time component of PWTT were analyzed. Recorded data suggest that the time component of PWTT may have a significant impact on the accuracy of estimating stroke volume during changes in SVR. However, further prospective clinical studies are required to test this hypothesis.