Technical performance and reflection capacity of the anaesthetic conserving device--a bench study with isoflurane and sevoflurane

麻醉剂保存装置的技术性能和反射能力——异氟烷和七氟烷的体外研究

阅读:1

Abstract

OBJECTIVE: The anaesthetic conserving device (AnaConDa), Sedana Medical, Sundbyberg, Sweden) facilitates administration of isoflurane or sevoflurane by liquid infusion. An anaesthetic reflector inside the device conserves exhaled anaesthetic and re-supplies it during inspiration. In this bench study, we examined the influence of infusion rates and ventilatory settings on the resulting anaesthetic concentrations on patient (C(pat)) and ventilator side of the reflector (C(loss)) to describe its technical performance. METHODS: A Puritan Bennett 840 ICU ventilator (Pleasanton, US), AnaConDa, and a test lung (3 l-chloroprene-bag) were assembled. Infusion rates (IR, 0.2-50 ml h(-1)), respiratory rates (RR, 5-40 breaths min(-1)), and tidal volumes (V(T), 0.3, 0.5, and 1.0 l) were varied. C(pat) was measured via a thin catheter in the middle of the 3 l-bag in steady state (online data storage and averaging over >10 min). C(loss) was calculated from IR (to yield the volume of vapour per unit of time), and expired minute volume (in which the vapour is diluted) on the assumption that, in the steady state, input by liquid infusion equals output through the reflector. RESULTS: At lower concentrations (C(pat) < 1 vol%) the ratio C(loss)/C(pat) was constant (R(C) = 0.096 +/- 0.012) for all combinations of IR, RR and V(T), both for isoflurane and sevoflurane. The device could efficiently reflect up to 10 ml vapour per breath (e.g. 2 vol% in 0.5 l). When exceeding this capacity, surplus vapour "spilled over" and R(C) markedly increased indicating decreased performance. CONCLUSIONS: The triple product minute volume times R(C) times C(pat) describes anaesthetic losses through the reflector. It can easily be calculated as long as the 10 ml reflection capacity is not exceeded and thus R(C) is constant. Increased minute ventilation necessitates increasing the IR to keep C(pat) constant. When using large V(T) and high C(pat) "spill over" occurs. This effect offers some protection against an inadvertent overdose.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。