Inaccuracy of Urine Output Measurements due to Urinary Retention in Catheterized Patients in the Burn ICU

烧伤重症监护室导尿患者因尿潴留导致尿量测量不准确

阅读:2

Abstract

Electronic urinary output monitors, intended to provide urine output information to guide fluid therapy during burn resuscitation, can be inaccurate because of airlocks causing urine retention in the drainage tube and bladder. In this study, the authors explore the effects of airlock formation on urine output measured using an electronic urinary output monitor connected to either a standard commercial drainage tubing system or a drainage tubing system with an automated airlock clearing mechanism. In a multicenter study in the burn intensive care unit, urine output was compared between 10 control patients with a standard commercial drainage tubing system and 10 test patients with a novel automated airlock clearing drainage tubing system. The comparison was focused on identifying the number and magnitude of surges in urinary output because of airlocks and associated periods of false oliguria. In the control group, 5 of 10 (50%) patients had drainage line flow impediments from 8 airlocks. In addition, control patients experienced six associated periods of false oliguria. Airlock surge volumes ranged from 50 to 329 ml, and false oliguria duration ranged from 39.4 to 185.2 minutes. In the test group, 0 of 10 (0%) patients had drainage line impediments from airlocks (P < .01), and hence, there were no periods of false oliguria. Airlocks and associated periods of false oliguria occur with standard commercial drainage tubing and are eliminated using an automated airlock clearing drainage tube. Electronic urinary output monitoring with self-clearing drainage has the potential to improve tracking of real-time urine output and decrease caregiver workload.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。