Accuracy of Different Topographic Instruments in Calculating Corneal Power after Myopic Photorefractive Keratectomy

不同地形图仪在近视准分子激光角膜切削术后计算角膜屈光力时的准确性

阅读:1

Abstract

PURPOSE: To compare the corneal power measurements obtained using different topographic instruments after myopic photorefractive keratectomy (PRK). METHODS: Patients with myopia who were candidates for corneal refractive surgery were sequentially included. Pre-PRK and six months post-PRK corneal powers were measured using Javal manual keratometer, Orbscan II, Galilei, Tomey TMS4, and EyeSys 2000 topographers. Measured values were compared with those obtained using the clinical history method (CHM). RESULTS: This study included 66 eyes of 33 patients. The lowest keratometric measurements were obtained using the Galilei topographer (42.98 ± 1.69 diopters, D) and the highest measurements were obtained using the Javal manual keratometer (43.96 ± 1.54 D) preoperatively. The same order was observed postoperatively. Effective refractive power (EffRP) measured using EyeSys was most similar to the values obtained using CHM (ICC, intraclass correlation coefficient = 0.951), followed by the total corneal power measured using the Galilei system (ICC = 0.943). The values obtained using the adjusted EffRP formula (EffRP - 0.015*Δ Refraction - 0.05) were more consistent with the values obtained using CHM (ICC = 0.954) compared to those obtained with the adjusted average central corneal power formula measured using the Tomey system (ICC = 0.919). CONCLUSION: Post-PRK corneal powers measured using the adjusted EffRP formula were the most similar to values obtained using CHM.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。