Central Corneal Thickness Measurement Using Ultrasonic Pachymetry, Rotating Scheimpflug Camera, and Scanning-slit Topography Exclusively in Thin Non-keratoconic Corneas

仅针对薄型非圆锥角膜,采用超声角膜测厚仪、旋转式 Scheimpflug 相机和扫描裂隙地形图进行中央角膜厚度测量

阅读:1

Abstract

PURPOSE: To evaluate the agreement among Pentacam, Orbscan and ultrasound (US) pachymetry for measurement of central corneal thickness (CCT) in thin corneas with normal topographic pattern. METHODS: We included 88 eyes of 44 refractive surgery candidates with thinnest pachymetric readings of 500 micrometers (μm) or less on Orbscan, a normal topographic pattern, no sign of keratoconus, and best corrected visual acuity (BCVA) of 20/20. Pentacam, Orbscan and US were performed in one session by the same examiner. Exclusion criteria were history of ocular surgery, topographic abnormalities suggesting forme fruste keratoconus or keratectasia, and recent contact lens wear. RESULTS: The difference in CCT measurements by US pachymetry and Orbscan II [using an acoustic factor (AF) of 0.92] ranged from -34 to +34 μm. The difference between the thinnest point and central readings measured by US reached 16 μm with Orbscan II (AF: 0.92) and 2 μm with Pentacam. Mean differences between the employed devices were 0.2 μm for Pentacam versus US (P = 0.727), 30.1 μm for uncorrected Orbscan versus US (P < 0.001), 10.4 μm for Orbscan II (AF = 0.92) versus US (P < 0.001), and 0.2 μm for Orbscan II (AF = 0.94) versus US (P = 0.851). CONCLUSION: In normal thin corneas, Pentacam demonstrated better agreement with US pachymetry as compared to corrected Orbscan readings. Results achieved by Orbscan were better consistent with US pachymetry using an AF of 0.94. We speculate that a dynamically graded AF in reverse proportion to CCT constitutes a better approach for correcting Orbscan measurements.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。