Validation of a data-driven motion-compensated PET brain image reconstruction algorithm in clinical patients using four radiotracers

利用四种放射性示踪剂在临床患者中验证数据驱动的运动补偿PET脑图像重建算法

阅读:1

Abstract

PURPOSE: Patients with dementia symptoms often struggle to limit movements during PET examinations, necessitating motion compensation in brain PET imaging to ensure the high image quality needed for diagnostic accuracy. This study validates a data-driven motion-compensated (MoCo) PET brain image reconstruction algorithm that corrects head motion by integrating the detected motion frames and their associated rigid body transformations into the iterative image reconstruction. Validation was conducted using phantom scans, healthy volunteers, and clinical patients using four radiotracers with distinct tracer activity distributions. METHODS: We conducted technical validation experiments of the algorithm using Hoffman brain phantom scans during a series of controlled movements, followed by two blinded reader studies assessing image quality between standard images and MoCo images in 38 clinical patients receiving dementia scans with [(18)F]Fluorodeoxyglucose, [(18)F]N-(3-iodopro-2E-enyl)-2beta-carbomethoxy-3beta-(4'-methylphenyl)-nortropane, [(18)F]flutemetamol, and a research group comprising 25 elderly subjects scanned with [(18)F]fluoroethoxybenzovesamicol. RESULTS: The Hoffman brain phantom study demonstrated the algorithm's capability to detect and correct for even minimal movements, 1-mm translations and 1⁰ rotations, applied to the phantom. Within the clinical cohort, where standard images were deemed suboptimal or non-diagnostic, all MoCo images were classified as having acceptable diagnostic quality. In the research cohort, MoCo images consistently matched or surpassed the standard image quality even in cases with minimal head movement, and the MoCo algorithm never led to degraded image quality. CONCLUSION: The PET brain MoCo reconstruction algorithm was robust and worked well for four different tracers with markedly different uptake patterns. Moco images markedly improved the image quality for patients who were unable to lie still during a PET examination and obviated the need for any repeat scans. Thus, the method was clinically feasible and has the potential for improving diagnostic accuracy.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。