First performance evaluation of easyPET.3D, a high-resolution and cost-effective benchtop preclinical PET scanner

对 easyPET.3D(一款高分辨率、高性价比的台式临床前 PET 扫描仪)进行首次性能评估

阅读:1

Abstract

PURPOSE: EasyPET.3D is a preclinical positron emission tomography (PET) scanner with a unique scanning method based on two face-to-face detector modules with two axes of motion. Its performance evaluation is presented using the NEMA NU-4 standards and an animal model. METHODS: Each detector module consists of a 32 × 2 pixelated cerium-doped lutetium-yttrium oxyorthosilicate (LYSO:Ce) scintillator array with individual crystals of 2.0 × 2.0 × 30 mm(3). The crystal arrays are coupled to 1.3 × 1.3 mm(2) silicon photomultipliers (SiPMs). The transaxial field-of-view (FoV) is adjustable up to 48 mm in diameter, and the axial length is 73 mm. The performance characterization includes spatial resolution, sensitivity, count rate, scatter fraction (SF), and image quality (IQ) measurements. Furthermore, mice experiments were conducted to evaluate the in vivo imaging capability. RESULTS: Spatial resolution at the FoV centre (CFoV) in radial, tangential, and axial directions was 0.98±0.08, 0.97±0.06 and 0.94±0.08 mm, respectively. An absolute sensitivity of 0.23% was measured at CFoV. The mouse-like phantom SF was 16% (913 cps at 18 MBq). Recovery coefficients in the IQ phantom varied from 21±34% to 85±50% (1 to 5 mm diameter rods, accordingly). The uniformity was 17.6%, and spill-over ratios for water-filled and air-filled chambers were 0.49 and 0.40, respectively. CONCLUSION: EasyPET.3D geometry favours the reduction of parallax error, despite its low sensitivity. The system linearity is suitable for the low range of activities (7-8 MBq) used for mice imaging. The overall performance showed that easyPET.3D has potential for entry-level preclinical applications.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。