Influence of X-ray computed tomography (CT) exposure and reconstruction parameters on positron emission tomography (PET) quantitation

X射线计算机断层扫描(CT)曝光和重建参数对正电子发射断层扫描(PET)定量的影响

阅读:2

Abstract

BACKGROUND: The CT of PET CT provides diagnostic information, anatomic localisation and attenuation correction (AC). When only AC is required, very lose dose CT is desirable. CT iterative reconstruction (IR) improves image quality with lower exposures however there is little data on very low dose IR CT for AC of PET. This work assesses the impact of CT exposure and reconstruction algorithm on PET voxel values. METHOD: An anthropomorphic torso phantom was filled with physiologically typical [18]F concentrations in heart, liver and background compartments. A 17-mm-diameter right lung "tumour" filled with [18]F was included (surrounding lung contained no 18[F]). PET was acquired followed by 24 CT acquisitions with varying CT exposures (15-50 mAs, 80-120 kVp, pitch 0.671 or 0.828). Each CT was reconstructed twice using filtered back projection (FBP) or IR and these used for AC of PET. The reference PET reconstruction (RR) used CT acquired at 50 mAs, 120 kVp, pitch 0.828, IR, all others were test PET reconstructions (TR). Regions of interest (ROIs) were drawn in the liver, soft tissue and over "tumour" on each TR and compared with the RR. Voxel values in each TR were compared to the RR using a paired t test and by calculating which and what proportion of voxels in each TR differed by a quantitatively significant difference (QSD) from the RR. RESULTS: TRs reconstructed using lower dose CTs underestimated mean and maximum ROI activity relative to the RR; greater with IR than FBP. Once CT dose index (CTDI) increased to 1 mGy, differences were less than QSD. On voxel analysis, all TRs were significantly different to the RR (p < 0.0001). TRs reconstructed at the lowest CT exposure with IR had 6% of voxels that differed by greater than QSD. Differences were reduced with increasing CTDI and FBP reconstruction. Voxels which exceeded the QSD were spatially localised to regions of high activity, interfaces between different attenuation and areas of CT beam hardening. CONCLUSIONS: Very low dose CT exposures are feasible for accurate PET AC. Scanner- and reconstruction-specific validation should be employed prior very low dose CT AC for PET.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。