Computational modelling and evaluation of cardiovascular response under pulsatile impeller pump support

脉动叶轮泵支撑下心血管反应的计算建模与评估

阅读:1

Abstract

This study presents a numerical simulation of cardiovascular response in the heart failure condition under the support of a Berlin Heart INCOR impeller pump-type ventricular assist device (VAD). The model is implemented using the CellML modelling language. To investigate the potential of using the Berlin Heart INCOR impeller pump to produce physiologically meaningful arterial pulse pressure within the various physiological constraints, a series of VAD-assisted cardiovascular cases are studied, in which the pulsation ratio and the phase shift of the VAD motion profile are systematically changed to observe the cardiovascular responses in each of the studied cases. An optimization process is proposed, including the introduction of a cost function to balance the importance of the characteristic cardiovascular variables. Based on this cost function it is found that a pulsation ratio of 0.35 combined with a phase shift of 200° produces the optimal cardiovascular response, giving rise to a maximal arterial pulse pressure of 12.6 mm Hg without inducing regurgitant pump flow while keeping other characteristic cardiovascular variables within appropriate physiological ranges.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。