Biomechanical evaluation of a new femoral stem design for total hip replacement in a canine model

在犬模型中对一种用于全髋关节置换的新型股骨柄设计进行生物力学评估

阅读:1

Abstract

PURPOSE: To evaluate the biomechanical properties of a novel total hip replacement femoral stem. METHODS: Eight pairs of femurs from dog cadavers were used. The femurs were separated into different groups. A novel femoral stem with a convex proximal portion (Stem B) was biomechanically evaluated and compared to awell-known veterinary collared stem (Stem A). Femoral stems were inserted into the contralateral femurs from the same dog, forming 16 constructs. A flexo-compression load was applied on the axial axis of each sample. Maximum strength, deflection, stiffness, and energy absorption were analysed. RESULTS: Group B constructs showed significantly higher values (p ? 0.05) for the variables, except stiffness. The mean maximum strength was 1,347 ± 357 N for Group A and 1,805 ± 123 N for Group B (p ? 0.0069). The mean deflection was5.54 ± 2.63 mm for Group A and 10.03 ± 3.99 mm for Group B (p ? 0.0056). For the energy variable, the force was 6,203 ± 3,488 N/mm for Group A and 12,885 ± 5,056 N/mm for Group B (p ? 0.0054). Stem B had greater maximum strength, deflection, and energy. CONCLUSIONS: The new stem was effective in neutralizing the impact of axial flexion-compression stresses during biomechanical tests in cadaveric models.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。