First clinical experience with a novel, mobile cone-beam CT system for treatment quality assurance in brachytherapy

新型移动式锥形束CT系统在近距离放射治疗治疗质量保证方面的首次临床应用

阅读:1

Abstract

BACKGROUND AND PURPOSE: On-site cone-beam computed tomography (CBCT) has gained in importance in adaptive brachytherapy during recent years. Besides treatment planning, there is increased need particularly for image-guidance during interventional procedures and for image-guided treatment quality assurance (QA). For this purpose, an innovative CBCT device was rolled out at our hospital as the first site worldwide. We present the first clinical images and experiences. MATERIALS AND METHODS: The novel CBCT system is constructed of a 121 cm diameter ring gantry, and features a 43.2 × 43.2 cm(2) flat-panel detector, wireless remote-control via tablet-PC, and battery-powered maneuverability. Within the first months of clinical operation, we performed CBCT-based treatment QA for a total of 26 patients (8 with breast, 16 with cervix, and 2 with vaginal cancer). CBCT scans were analyzed regarding potential movements of implanted applicators in-situ during the brachytherapy course. RESULTS: With the presented device, treatment QA was feasible for the majority of patients. The CBCT scans of breast patients showed sufficient contrast between implanted catheters and tissue. For gynecologic patients, a distinct visualization of applicators was achieved in general. However, reasonable differentiations of organic soft tissues were not feasible. CONCLUSION: The CBCT system allowed basic treatment QA measures for breast and gynecologic patients. For image-guidance during interventional brachytherapy procedures, the current image quality is not adequate. Substantial performance enhancements are required for intraoperative image-guidance.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。