Evaluation of prostheses retained zygomatic and dental implants in large defects in the maxilla due to tumors or major trauma by biomechanical 3- dimensional finite element analysis

利用生物力学三维有限元分析评估上颌骨因肿瘤或严重创伤造成的大缺损中颧骨和种植牙固位体的修复效果

阅读:2

Abstract

BACKGROUND: Zygomatic implants are becoming an ideal treatment approach for implant-supported prosthesis treatment developed for the atrophic maxilla. This study aims to evaluate the amount and distribution of stress in implants and peri-implant bone using different implant-supported prosthesis configurations in Aramany Class I maxillary defects through 3-dimensional finite element analysis. METHODS: A 3-dimensional finite element model of the Aramany class I defect was created. Three different implant-supported prostheses were modelled: model 1: 1 zygomatic implant and 3 dental implants, model 2: 1 zygomatic implant and 2 dental implants, and model 3: 2 zygomatic implants. Vertical and horizontal loads of 150 N were applied in 6 different ways to the defected and non-defective areas. Maximum principal stress and von mises stresses in the bone surrounding the implants were evaluated. RESULTS: When all loading conditions were evaluated with both porcelain layer on Co-Cr framework and acrylic layer on acrylic framework, the highest maximum principal stress value was observed in Model 3. In contrast, the lowest value was observed in Model 1. The highest maximum principal stress occurred when a horizontal load was applied simultaneously to both the defective and non-defective areas. In contrast, the lowest value was observed when a vertical load was applied to the non-defective area. The von Mises stress values were found to be similar across all models when both restoration materials were used. CONCLUSIONS: Based on the results of this study, it can be concluded that increasing the number of implants in the non-defective area reduces the highest stress value while using acrylic as a restoration material slightly increases the stress value.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。