Continuous Noninvasive Remote Automated Blood Pressure Monitoring With Novel Wearable Technology: A Preliminary Validation Study

基于新型可穿戴技术的连续无创远程自动血压监测:初步验证研究

阅读:1

Abstract

BACKGROUND: Wearable continuous monitoring biosensor technologies have the potential to transform postoperative care with early detection of impending clinical deterioration. OBJECTIVE: Our aim was to validate the accuracy of Cloud DX Vitaliti continuous vital signs monitor (CVSM) continuous noninvasive blood pressure (cNIBP) measurements in postsurgical patients. A secondary aim was to examine user acceptance of the Vitaliti CVSM with respect to comfort, ease of application, sustainability of positioning, and aesthetics. METHODS: Included participants were ≥18 years old and recovering from surgery in a cardiac intensive care unit (ICU). We targeted a maximum recruitment of 80 participants for verification and acceptance testing. We also oversampled to minimize the effect of unforeseen interruptions and other challenges to the study. Validation procedures were according to the International Standards Organization (ISO) 81060-2:2018 standards for wearable, cuffless blood pressure (BP) measuring devices. Baseline BP was determined from the gold-standard ICU arterial catheter. The Vitaliti CVSM was calibrated against the reference arterial catheter. In static (seated in bed) and supine positions, 3 cNIBP measurements, each 30 seconds, were taken for each patient with the Vitaliti CVSM and an invasive arterial catheter. At the conclusion of each test session, captured cNIBP measurements were extracted using MediCollector BEDSIDE data extraction software, and Vitaliti CVSM measurements were extracted to a secure laptop through a cable connection. The errors of these determinations were calculated. Participants were interviewed about device acceptability. RESULTS: The validation analysis included data for 20 patients. The average times from calibration to first measurement in the static position and to first measurement in the supine position were 133.85 seconds (2 minutes 14 seconds) and 535.15 seconds (8 minutes 55 seconds), respectively. The overall mean errors of determination for the static position were -0.621 (SD 4.640) mm Hg for systolic blood pressure (SBP) and 0.457 (SD 1.675) mm Hg for diastolic blood pressure (DBP). Errors of determination were slightly higher for the supine position, at 2.722 (SD 5.207) mm Hg for SBP and 2.650 (SD 3.221) mm Hg for DBP. The majority rated the Vitaliti CVSM as comfortable. This study was limited to evaluation of the device during a very short validation period after calibration (ie, that commenced within 2 minutes after calibration and lasted for a short duration of time). CONCLUSIONS: We found that the Cloud DX's Vitaliti CVSM demonstrated cNIBP measurement in compliance with ISO 81060-2:2018 standards in the context of evaluation that commenced within 2 minutes of device calibration; this device was also well-received by patients in a postsurgical ICU setting. Future studies will examine the accuracy of the Vitaliti CVSM in ambulatory contexts, with attention to assessment over a longer duration and the impact of excessive patient motion on data artifacts and signal quality. TRIAL REGISTRATION: ClinicalTrials.gov NCT03493867; https://clinicaltrials.gov/ct2/show/NCT03493867.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。