Long-term outcomes and durability of balloon-expandable TAVI in small and large annuli

小瓣环和大瓣环球囊扩张式经导管主动脉瓣置换术的长期疗效和耐久性

阅读:2

Abstract

BACKGROUND: Evidence regarding the long-term outcomes and durability of balloon-expandable transcatheter heart valves (BE-THVs) after transcatheter aortic valve implantation (TAVI) is still scarce. AIMS: This study evaluates these outcomes and further examines their association with the annular size in patients. METHODS: A total of 2,699 patients who had undergone TAVI with a BE-THV at least 5 years prior to our study were analysed. A small aortic annulus (SAA) was defined as an area ≤430 mm(2); any larger annuli were labelled large aortic annuli (LAA). The primary endpoints were the incidence of all-cause mortality and bioprosthetic valve failure (BVF) between the SAA and LAA groups. As a subanalysis, the primary endpoints were examined in relation to postprocedural mean pressure gradient (mPG) ≥20 mmHg, severe prosthesis-patient mismatch (PPM), and sex differences, comparing SAA and LAA each time. RESULTS: Overall, 66.4% (n=1,793) of patients were categorised into the SAA group. At 7 years after TAVI, the cumulative all-cause mortality showed differences between the SAA and LAA groups (55.2% vs 58.6%), while BVF assessed by Gray's test was similar between the groups (3.3% vs 2.7%). The Cox multivariable analysis revealed no association between SAA and worse prognosis (hazard ratio 1.07, 95% confidence interval: 0.85-1.36; p=0.56). There were no significant differences in mortality or BVF regarding an mPG ≥20 mmHg, severe PPM, or sex between the SAA and LAA groups (allp>0.05). CONCLUSIONS: Annular size differences were not found to influence long-term outcomes or valve durability following TAVI with a BE-THV, suggesting that other factors warrant further investigation.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。