Extended Intraoperative Longitudinal 3-Dimensional Cone Beam Computed Tomography Imaging With a Continuous Multi-Turn Reverse Helical Scan

采用连续多圈反向螺旋扫描的扩展术中纵向三维锥形束计算机断层扫描成像

阅读:2

Abstract

OBJECTIVES: Cone beam computed tomography (CBCT) imaging is becoming an indispensable intraoperative tool; however, the current field of view prevents visualization of long anatomical sites, limiting clinical utility. Here, we demonstrate the longitudinal extension of the intraoperative CBCT field of view using a multi-turn reverse helical scan and assess potential clinical utility in interventional procedures. MATERIALS AND METHODS: A fixed-room robotic CBCT imaging system, with additional real-time control, was used to implement a multi-turn reverse helical scan. The scan consists of C-arm rotation, through a series of clockwise and anticlockwise rotations, combined with simultaneous programmed table translation. The motion properties and geometric accuracy of the multi-turn reverse helical imaging trajectory were examined using a simple geometric phantom. To assess potential clinical utility, a pedicle screw posterior fixation procedure in the thoracic spine from T1 to T12 was performed on an ovine cadaver. The multi-turn reverse helical scan was used to provide postoperative assessment of the screw insertion via cortical breach grading and mean screw angle error measurements (axial and sagittal) from 2 observers. For all screw angle measurements, the intraclass correlation coefficient was calculated to determine observer reliability. RESULTS: The multi-turn reverse helical scans took 100 seconds to complete and increased the longitudinal coverage by 370% from 17 cm to 80 cm. Geometric accuracy was examined by comparing the measured to actual dimensions (0.2 ± 0.1 mm) and angles (0.2 ± 0.1 degrees) of a simple geometric phantom, indicating that the multi-turn reverse helical scan provided submillimeter and degree accuracy with no distortion. During the pedicle screw procedure in an ovine cadaver, the multi-turn reverse helical scan identified 4 cortical breaches, confirmed via the postoperative CT scan. Directly comparing the screw insertion angles (n = 22) measured in the postoperative multi-turn reverse helical and CT scans revealed an average difference of 3.3 ± 2.6 degrees in axial angle and 1.9 ± 1.5 degrees in the sagittal angle from 2 expert observers. The intraclass correlation coefficient was above 0.900 for all measurements (axial and sagittal) across all scan types (conventional CT, multi-turn reverse helical, and conventional CBCT), indicating excellent reliability between observers. CONCLUSIONS: Extended longitudinal field-of-view intraoperative 3-dimensional imaging with a multi-turn reverse helical scan is feasible on a clinical robotic CBCT imaging system, enabling long anatomical sites to be visualized in a single image, including in the presence of metal hardware.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。