High-permittivity thin dielectric padding improves fresh blood imaging of femoral arteries at 3 T

高介电常数薄介电垫片可改善3T磁场下股动脉新鲜血液的成像效果。

阅读:1

Abstract

OBJECTIVES: Fresh blood imaging (FBI) is a useful noncontrast magnetic resonance angiographic (MRA) method for the assessment of peripheral arterial disease, particularly for imaging patients with poor renal function. Compared with 1.5 T, 3 T enables higher signal-to-noise ratio and/or spatiotemporal resolution in FBI. Indeed, previous studies have reported successful FBI of the calf station at 3 T. However, FBI of the thigh station at 3 T has been reported to suffer from signal void in the common femoral artery of 1 thigh only because of the radial symmetry in transmit radiofrequency field (B1+) variation. We sought to increase the signal of femoral artery in FBI at 3 T using high-permittivity dielectric padding. MATERIALS AND METHODS: We performed FBI and B1+ mapping of the thigh station at 3 T in 13 human subjects to compare the following 3 dielectric padding settings: no padding, commercially available thick (approximately 5 cm) dielectric padding, and high-permittivity thin (approximately 2 cm) dielectric padding. We characterized the radial symmetry in B1+ variation as well as its impact on the FBI signal at baseline and how dielectric padding improves B1+ and FBI. We evaluated the quality of 3 FBI MRA acquisitions using quantitative (ie, contrast-to-noise ratio of femoral arteries) and qualitative (ie, conspicuity of femoral arteries) analyses. RESULTS: With the subjects positioned on the magnetic resonance table in feet-first, supine orientation, the radial symmetry in B1+ variation attenuates the signal in the right common femoral artery. The signal void can be improved partially with commercial padding and improved further with high-permittivity padding. Averaging the results over the 13 subjects, the mean B1+, contrast-to-noise ratio, and conspicuity scores for the right common femoral artery were significantly higher with high-permittivity padding than with commercial padding and baseline (P < 0.001). CONCLUSIONS: Our study shows that high-permittivity dielectric padding can be used to increase the signal of femoral artery in FBI at 3 T.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。