Comparison of the accuracy of an ultrasonic-based jaw tracking device with conventional electronic tracking device

超声波颌部跟踪装置与传统电子跟踪装置的精度比较

阅读:1

Abstract

PURPOSE: This study aimed to evaluate the accuracy of the ultrasonic jaw tracking system by comparing with the conventional electronic system in recording condylar movements. MATERIALS AND METHODS: Twenty-six subjects with normal occlusion participated in the study. The CADIAX(®) 4 and Jaw Motion Analyzer (JMA) systems were used to record condylar movement trajectories during mandibular border movements (protrusive/retrusive, lateral, and wide mouth opening), with each movement repeated three times. Both systems used facebows and sensors to locate the condylar axis points and capture movement trajectory data. Paired t-tests were used for normally distributed data, while the Wilcoxon rank-sum test was applied to non-normally distributed data. The level of significance was set at α = .05. RESULTS: The maximum condylar displacement in the sagittal plane during mandibular border movements and the sagittal condylar inclination (SCI) values on both the left and right sides showed no significant difference between the two systems (P > .05). The Bennett angle (BA) values on both the left and right sides measured by the JMA system were significantly higher than those measured by the CADIAX(®) 4 system (P < .05). The comfort levels of the JMA system were significantly higher than the CADIAX(®) 4 system (P < .05). CONCLUSION: Through this study, it was found that the accuracy of the ultrasonic jaw tracking system was comparable with the conventional electronic system, except for the Bennett angle measurement. In terms of comfort and ease of use, the ultrasonic jaw tracking system is more favored.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。