The Effect of a Novel Transcatheter Edge-to-Edge Mitral Valve Repair Device in a Porcine Model of Mitral Regurgitation

新型经导管缘对缘二尖瓣修复装置在猪二尖瓣反流模型中的作用

阅读:1

Abstract

BACKGROUND: A new technique has been developed for treating mitral regurgitation (MR) via a transapical approach, which encompasses an easy-to-use leaflet clamp and a smaller-sized delivery system (14F-16F). OBJECTIVES: We aimed to evaluate the effectiveness of this device in a porcine model of acute MR. METHODS: Acute MR was induced in 36 anesthetized porcine subjects by severing the chordae supporting the corresponding segment of the leaflet. The ValveClamp system was then transapically implanted on the prolapsing segment under epicardial echocardiographic guidance. Echocardiographic assessments were performed before and after the transapical interventions. All of the animals were killed 30 days after the procedure to verify the proper location of the implanted devices. RESULTS: Epicardial echocardiography revealed severe MR (n = 26) or moderate to severe MR (n = 10) in the pig model of acute MR. Overt MR reduction was observed following the procedure through echocardiography; residual MR was absent in 10 cases, mild in 17 cases, and moderate in 9 cases. There was no evidence of mitral stenosis at the end of the procedure in terms of mitral valve area and mitral valve pressure gradient. Autopsy demonstrated that all ValveClamp devices were precisely placed to clamp the prolapsing segment of the mitral valve. CONCLUSIONS: Transapical implantation of the ValveClamp device under epicardial echocardiographic guidance was effective and safe in reducing acutely induced MR in this pig model. It is potentially applicable as a novel user-friendly transcatheter edge-to-edge mitral valve repair device for the treatment of MR in humans.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。