A NEW ANATOMICAL PLATE FOR EXTRA-ARTICULAR DISTAL HUMERAL FRACTURES: BIOMECHANICAL STUDY

一种用于关节外远端肱骨骨折的新型解剖钢板:生物力学研究

阅读:1

Abstract

INTRODUCTION: We compared the mechanical properties of two fixation techniques for the treatment of extra-articular distal third humeral fractures. MATERIALS AND METHODS: Two groups were created from twenty-four humeri. Group 1 was instrumented using a new, precontoured, 8-hole (3.5-mm-diameter) locking compression plate (LCP) placed anterolaterally. Group 2 was instrumented using an 8-hole (3.5-mm-diameter) precontoured posterolateral LCP plate placed on the distal humerus. Four-point bending tests and torsion tests were performed until the specimens broke. RESULTS: The four-point bending stiffness test showed that the stiffness of anterolaterally fixed humeri was significantly higher than that of posterolaterally fixed humeri (p<0.05). Torsion testing revealed that posterolateral fixation was associated with better yield strength (p<0.05), but the torsional stiffness did not differ significantly between the two plates (p> 0.05). CONCLUSIONS: The anterolateral plate exhibited higher bending stiffness and torsional yield strength than the posterolateral plate. Anterolateral plate fixation can thus be used to manage extra-articular distal humeral fractures. Multiaxial locking screws ensure rigid fixation, allow early elbow motion without olecranon fossa impingement, and prevent iatrogenic injury of the triceps muscle. Level of Evidence I, Therapeutic Studies Investigating the Results of Treatment.