Feasibility of Parylene C for encapsulating piezoelectric actuators in active medical implants

使用 Parylene C 封装有源医疗植入物中的压电执行器的可行性

阅读:8
作者:Alina Kohler #, Felix Blendinger #, Sonja Müller, Ulrich Mescheder, Volker Bucher

Abstract

Parylene C is well-known as an encapsulation material for medical implants. Within the approach of miniaturization and automatization of a bone distractor, piezoelectric actuators were encapsulated with Parylene C. The stretchability of the polymer was investigated with respect to the encapsulation functionality of piezoelectric chips. We determined a linear yield strain of 1% of approximately 12-µm-thick Parylene C foil. Parylene C encapsulation withstands the mechanical stress of a minimum of 5×105 duty cycles by continuous actuation. The experiments demonstrate that elongation of the encapsulation on piezoelectric actuators and thus the elongation of Parylene C up to 0.8 mm are feasible.