Thickness of orthodontic clear aligners after thermoforming and after 10 days of intraoral exposure: a prospective clinical study

正畸透明矫正器热成型后及口内暴露10天后的厚度:一项前瞻性临床研究

阅读:1

Abstract

BACKGROUND: Clear aligners (CA) are among the most chosen orthodontic therapies for patients who require an invisible treatment. Previous studies showed that the thermoforming process and the complexity of the intraoral environment might alter the properties of these devices. The aim of the current prospective clinical study was to assess the thickness changes of the CA after 10 days of intraoral use. The secondary aim was to assess the reproducibility of the thermoforming process, in terms of aligner thickness. MATERIALS AND METHODS: CA from 18 consecutive patients (13 women, 5 men, mean age 28.8 ± 9.6 years) were investigated. Before intraoral exposure (T0), the thickness of the unused CA was measured at different occlusal points on a 3D model with a dedicated software (Geomagic Qualify 2013; 3D Systems, Rock Hill, SC, USA). Two CA configurations were studied: passive maxillary aligner (P-no tooth movement; no shape for attachments) and active maxillary aligner (A-tooth movement; shape for attachments and divot). The used aligners were returned after 10 days (T1) and the thickness measurements were repeated. A Student's t test for paired data (T1 vs. T0) was applied to compare the thicknesses of used and unused devices (significance level after Bonferroni correction for multiple comparison was set at p < 0.0014). Furthermore, to study the reproducibility of the thermoforming process, P and A aligners were thermoformed twice, and the thicknesses of the two unused thermoformed devices were compared by means of Student's t test for paired data (significance level after Bonferroni correction for multiple comparison was set at p < 0.0014) and Dahlberg's error. RESULTS: The thermoforming process showed good reproducibility for both aligner configurations, with a maximum Dahlberg's error of 0.13 mm. After intraoral use, the thickness of P showed some statistically significant, but not clinically relevant, thickness changes as compared to the unused aligners, while A did not show any significant changes. CONCLUSION: Considering the thickness changes, the thermoforming process is reliable both with active and passive aligner configurations. Also, the CA examined show good thickness stability after physiological intraoral ageing in a population of healthy adults.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。